Recall Z-0219-2015 ยท HSB Rod, Fixation, Intramedullary And Accessories

Synthes, Inc.

10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM - Sterile: intended to treat stable and unstable fractures of the proximal femur

DateNov 14, 2014 (initiated Oct 29, 2014)
StatusTerminated
ReasonThe 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM (Sterile) was assembled incorrectly. If the non conformance goes undetected, it may impact the locking function post-operatively and compromise the bone reduction and construct stability which may lead to non-union/malunion.
Root causeEmployee error
ActionAn urgent medical device recall notice, dated October 29, 2014, was sent to end users that identified the product, problem and action to be taken. Customers were instructed to return the device to Synthes along with completing the verification section of the notice.
Quantity6
DistributionUS Distribution in the states of: AK, FL, OH, OK, and TX.