Recall Z-0506-2015 ยท HQC Unit, Phacofragmentation

Alcon Research, Ltd.

INFINITI VISION SYSTEM ULTRASOUND FMS. Single-use fluid management system. STERILE. Model Numbers: 8065741080, 8065741081, 8065741082, 8065741083, 8065741085, 8065741087, 8065741088, 8065741089, 8065741093, 8065741096, 8065741097, 8065741099, 8065750266, 8065750268, 8065750274, 8065750278, 8065750280, 8065750281, 8065751009, 8065751010, 8065751011, 8065751716, 8065751717, 8065751718, 806575171

DateDec 4, 2014 (initiated Oct 30, 2014)
StatusTerminated
ReasonAlcon is conducting a medical device recall due to the possible presence of pieces of tubing in the aspiration line.
Root causeProcess control
ActionThe firm, Alcon, sent an "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letter dated October 30, 2014 to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately stop further use of the affected INFINITI FMS; review your inventory to determine if you have any affected units; segregate the potentially-affected product to ensure it is not used; complete and return the attached Response Form via fax: 817-302-4337 or email: Market.Actions@Alcon.com to Alcon, even if you have zero(0) units in inventory; immediately forward to all depa
Quantity147,473 units
DistributionWorldwide Distribution: US (nationwide) and countries of: Canada, Mexico, Australia, Belgium, Hong Kong, Japan, Bosnia, France, Estonia, Germany, Greece, Hungary, Iceland, Israel, Iraq, Latvia, Saudi Arabia, Spain, UAE, Paraguay, Pakistan, Vietnam, and Kurdistan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120912INFINITI(R) VISION SYSTEM
Alcon Manufacturing, Ltd., Jun 15, 2012