Recall Z-0996-2015 ยท GBB Chair, Surgical, Ac-Powered

TransMotion Medical Inc

Powered stretcher chair used during facial and eye surgeries. One chair per package.

DateJan 16, 2015 (initiated Nov 14, 2014)
StatusTerminated
ReasonThe firm is finding that the stretcher chair controllers are moving the chairs involuntarily.
Root causeComponent design/selection
ActionTransMotion Medical Inc. sent an Urgent Medical Device Recall letter dated November 4, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: Discontinue use of the device and contact TransMotion Medical Inc. for a replacement. Suspected faulty pendants must be destroyed by severing the cord and discarding the control. Replacement of the pendant is a long-term solution to the issue due to replacement pendants being not of the same type or design as the failed pendant. Accompanying the reca
Quantity1,127 units
DistributionWorldwide Distribution - US Distribution and the countries including Canada, Australia, Chile, China, Costa Rica, Germany, Algeria, Israel, Italy, South Korea, Malaysia, Nigeria, Netherlands, Poland, Portugal, Saudi Arabia, Singapore, Thailand, and Venezuela.