Recall Z-0579-2015 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ArthroCare Medical Corporation

Model Number ASHA4730-01: Ambient HipVac 50 Wand with Integrated Finger Switches and Model Number ASHA5050-01: Ambient KVac Wand with Integrated Finger Switches. The Wands are indicated for the resection, ablation, and coagulation of soft tissue, and hemostasis of blood vessels in arthroscopic and orthopedic procedures.

DateDec 16, 2014 (initiated Oct 27, 2014)
StatusTerminated
ReasonPotential breach of sterile barrier due to defective product tray.
Root causePackage design/selection
ActionThe firm notified their consignees on 10/29/14 via letter with a response form included. The letter requested the consignees discontinue use of the device and return it to the vendor.
Quantity23,119 devices
DistributionWorldwide Distribution-US (nationwide) and the countries of Austria, Australia, Belgium, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Ireland, Malaysia, Netherlands, Norway, Poland, Spain, South Africa, Sweden, Switzerland, Turkey, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083306MODIFICATION TO ARTHROCARE ARTHROWANDS
Arthrocare Corp., Dec 10, 2008