ClearanceBrief
Recall Z-0841-2015 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

GE OEC Medical Systems, Inc

GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.

DateJan 13, 2015 (initiated Sep 15, 2014)
StatusTerminated
ReasonmA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Root causeRadiation Control for Health and Safety Act
ActionPlanned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8.
Quantity61 units installed in US
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123603OEC BRIVO 865 ADVANCE, OEC BRIVO 785 ESSENTIAL, AND OEC BRIVO 715 PRIME
Ge Healthcare Surgery GE Oec Medical Systems, Inc., May 17, 2013