Recall Z-0840-2015 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
GE OEC Medical Systems, Inc
GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
| Date | Jan 13, 2015 (initiated Sep 15, 2014) |
| Status | Terminated |
| Reason | mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3). |
| Root cause | Radiation Control for Health and Safety Act |
| Action | Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8. |
| Quantity | 61 units installed in US |
| Distribution | Worldwide Distribution |