Recall Z-0235-2015 ยท KPQ System, Simulation, Radiation Therapy

Siemens Medical Solutions USA, Inc

Siemens SimView NT and SimView NT with IDI simulation system components: Product Usage: The intended use of the Radiation Therapy Simulation system is used for simulation of treatment planning prior to the therapeutic treatment of cancer.

DateNov 20, 2014 (initiated Oct 10, 2014)
StatusTerminated
ReasonSiemens became aware of a very rare occurrence where an intended movement of the SimView NT simulation system components could not be stopped with usual hand control commands or by use of the collision detection functionality. It is possible that simulation system components could collide with patient and/or system equipment.
Root causeDevice Design
ActionCustomers were initially notified about the issue by letter on January 7, 2014 (this action was reported as Z-0786-2014 which has been terminated). Siemens promised to notify them when a fix became available. Customer letter, dated November 7, 2014, was sent to notify end users that a fix has become available and Siemens would contact them to schedule an appointment for the correction.
Quantity6
DistributionWorldwide Distribution - US Nationwide in the states of PA, KY, LA, MA, MI, and CA; and countries of: Argentina, Belarus, Bulgaria, Costa Rico, P.R. China, Curacao, Egypt, Repulic Korea, Germany, Greece, Hungary, Italy, Iraq, Jordan, Malaysia, Palestanian Ter, Poland, Thailand, Turkey, and Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953233SIMVIEW 3000 CT OPTION
Siemens Medical Solutions USA, Inc., Nov 29, 1996