Recall Z-0226-2015 ยท NKB Thoracolumbosacral Pedicle Screw System

Synthes (USA) Products LLC

Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System. Intended for posterior pedicle screw fixation (Tl OS2- ilium), posterior hook fixation (Tl-L5), or anterolateral fixation (T8-L5).

DateNov 26, 2014 (initiated Oct 23, 2014)
StatusTerminated
ReasonTwo nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct specification.
Root causeProcess control
ActionDePuy Synthes sent an Urgent Notice: Medical Device Recall letter, dated October 23, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check inventory and remove affected stock, call Synthes to obtain a Return Authorization for the device, and to return the completed verification section of the response form by fax to (610) 430-7083 or Scan/email: Fieldaction@synthes.com. Customers with questions were instructed to call 610-719-5450.
Quantity24
DistributionDistributed to MI, MT, and CO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100952MODIFICATION TO SYNTHES MATRIX SYSTEM
Synthes Spine, Aug 5, 2010