Recall Z-0184-2015 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

GE Healthcare Infinia (including Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye, Infinia II Hawkeye 4, Infinia II VC, Infinia II VC Hawkeye, and Infinia II VC Hawkeye 4) Models: H3000WC, H3000WD, H3000WE, H3000WM, H3000WN, H3000WP, H3000WR, H3000WT, H3000WW, H3000WY, H3000WZ, H3000YF, H3000YG, H3000YM, H3000YS, H2400YJ, H2400YK. Hawkeye 4 Option For Dual-Head Variable Angle Gamma Ca

DateDec 16, 2014 (initiated Oct 14, 2014)
StatusTerminated
ReasonGE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.
Root causeNonconforming Material/Component
ActionThe firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter, GEHC Ref# 40860, dated October 14, 2014 to its consignees. The letter described the Safety Issue, safety Instructions, Affected Product Details, Product Correction, actions to be taken and Contact Information. For questions, consignees in the United States, can contact 1-800-437-1171, for other countries they can contact their GE Healthcare Service Representative.
Quantity2479
DistributionWorldwide Distribution: US (nationwide) including Guam, Puerto Rico and DC; and countries of: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BELRARUS, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FINLAND, F
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022960GE QUASAR NUCLEAR MEDICINE SYSTEM
Ge Medical Systems F.I. Haifa, Sep 19, 2002
K052434HAWKEYE 4 OPTION FOR DUAL-HEAD VARIABLE ANGLE GAMMA CAMERA
GE Medical Systems, Sep 27, 2005
K991841HAWKEYE OPTION FOR DUAL-HEAD VARIABLE ANGLE GAMMA CAMERA
Ge Medical Systems F.I. Haifa, Aug 26, 1999