ClearanceBrief
Recall Z-0993-2015 ยท OUG Medical Device Data System

Maquet Medical Systems USA

TEGRIS System manufactured by MAQUET GMBH in Germany The Maquet Tegris OR Integration System is designed to be used as the central operating unit in an operating room. The integration system has two main functions: Recording and distribution of images and videos and interaction with medical and non-medical devices.

DateJan 16, 2015 (initiated Aug 21, 2014)
StatusTerminated
ReasonThe wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.
Root causeSoftware design (manufacturing process)
ActionMaquet sent an Urgent Medical Device Field Correction Letter dated August 21, 2014, to all affected customers via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take the following measures until the issue is corrected: The TEGRIS equipped with a software module version up to 2.1.X.X should not be used in conjunction wiht a MAGNUS operating table system. All other MAQUET operating tables are not affected. A MAQUET service representative would contact them and arrange to have their TEGRIS software mod
Quantity6 units
DistributionUS Nationwide Distribution including NE, NY and PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.