Recall Z-0463-2015 ยท JJH Clinical Sample Concentrator

Abbott Molecular

Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.

DateNov 25, 2014 (initiated Oct 23, 2014)
StatusTerminated
ReasonSome versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.
Root causeSoftware design (manufacturing process)
ActionField Correction Recall / Urgent Field Safety Notice letters dated September 23, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to: 1) review the app spec files that are installed on the m2000sp system and delete any incompatible app specs; 2) if one or more app spec files were deleted, installation of new app spec file(s) is needed; and, 3) review this information with laboratory personnel and retain the letter for future reference. Customers with questions about this market action can contact Technical Support at 800-553-7042, Option 2, or
Quantity9 CD ROMs (US); 1,273 CD ROMs (Foreign)
DistributionWorldwide distribution. US nationwide (CA, CT, FL, NC, PA, SC, and DC), Algeria , Australia , Austria , Belgium , Bosnia and Herzegovina , Cameroon , Canada , Chad , Colombia , Costa Rica , Croatia , Denmark , Egypt , Ethiopia , Finland , France , Germany , Greece , Guinea , Honduras , Hong Kong , I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092705ABBOTT M2000SP AND ABBOTT M2000RT, MODELS 9K14-01 (G-SERIES), 9K14-02 (E-SERIES), 9K1501
Abbott Molecular, Inc., May 28, 2010