Recall Z-0170-2015 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Siemens Healthcare Diagnostics, Inc.

Rapid Neg BP Combo Panel Type 3; Catalog number B 1017-117. MicroScan Rapid panels are designed for use in determining identification to the species level and/or antimicrobial agent susceptibility of rapidly growing aerobic and facultatively anaerobic gram negative bacilli.

DateNov 7, 2014 (initiated Oct 17, 2014)
StatusTerminated
ReasonAn increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Neg Urine Combo Panel Type 1 used with MicroScan Microbiology Systems.
Root causeNonconforming Material/Component
ActionCustomer Recall notification letters were sent on October 17, 2014. Letters recommend that any remaining inventory be discarded and that consideration be given to reviewing previous test result, conduct patient follow up and/or repeat testing if isolates are still available. A second letter clarifying the recall was issued November 7, 2014 which provided additional information as requested by CDRH.
Quantity1381
DistributionDistributed US (nationwide) including the states of IA, CA, NY, WI, MI, SD, NC, FL, HI, OH, WA, TX, WA, IN, NM, MS, KY, IL, UT, ND, and WY and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.