Recall Z-0222-2015 ยท HQF Laser, Ophthalmic

Nidek Inc

MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic device used for retinal photocoagulation for treatment of ocular fundus diseases.

DateNov 17, 2014 (initiated May 24, 2013)
StatusTerminated
ReasonRetrospective review found that a Engineering Change Order initiated in May 2013 for a memory function problem was not reported as required.
Root causeSoftware design
ActionNidek Engineers visited all affected sites to install new software version. For questions regarding this recall call 510-353-7785.
Quantity5
DistributionUS Distribution in the states of: CA, NE, and NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110228NIDEK MULTICOLOR LASER PHOTOCOAGULATOR
Nidek Incorporated, Feb 23, 2011