Recall Z-0223-2015 ยท LZS Excimer Laser System

Nidek Inc

Nidek Excimer Laser Corneal Surgery System, model EC-5000 CXIII. Opthalmic Device. The system is composed as a laser generator which produces an excimer laser radiation of wavelength 193nm, a beam delivery unit, an optical system for observation, a gas system and a computer for system control.

DateNov 17, 2014 (initiated Aug 6, 2012)
StatusTerminated
ReasonContact failure in F1 and F4 fuse holders was detected on the EC-5000 CXIII Excimer Laser Corneal Surgery System that could cause a serious adverse event for the patient.
Root causeComponent design/selection
ActionNidek Engineers visited each site to perform correction - fuse holders have been inspected and replaced if necessary. An Engineering Change Order was issued on June 1, 2012, for inspection and/or replacement of affected fuse holders by Nidek Service Engineers. For further questions please call (510) 353-7785.
Quantity14 units
DistributionUS Distribution including the states of CO, NY, MO, LA, ME, CA, MI, AZ, CT and FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.