Recall Z-0421-2015 ยท FRN Pump, Infusion
Hospira Inc.
GemStar infusion pump, list numbers 13000 and 13100, GemStar Infusion Pump (pain management) list number 13150. Product Usage The pump is suitable for intravenous, arterial, subcutaneous, short-term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products. The epidural route is recommended to provide anesthesia or administe
| Date | Nov 24, 2014 (initiated Mar 19, 2013) |
| Status | Terminated |
| Reason | Some patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients are obtaining access to these codes via website sponsors by downloading the GemStar User/Service Manual featured on the sponsors' websites. The GemStar User/Service Manual should be removed from these w |
| Root cause | Other |
| Action | Hospira sent a letters dated March 19, 2013 to Internet website sponsors to request removal of the GemStar User/Service Manual from their websites as soon as possible. The Internet website sponsors were requested to return a confirmation form after removing the GemStar User Service Manual from their site. For questions call 224-212-2891. |
| Quantity | Not Applicable |
| Distribution | Worldwide Distribution - US: Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Croatia, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Icelanofd, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Netherlands, New Zealand, Oman, Portugal, Qatar, Rom |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K060806 | HOSPIRA GEMSTAR INFUSION PUMP SYSTEM, 7 THERAPY, MODEL 13087; 6 THERAPY, MODEL 13086; PAIN MANAGEMENT, MODEL 13088 Hospira, Inc., Apr 21, 2006 |