Recall Z-0202-2015 ยท HSB Rod, Fixation, Intramedullary And Accessories

Medacta Usa

GMK Intramedullary Extension Rod 100 mm is a rod to obtain correct leg alignment during total knee replacement. Rod to obtain correct leg alignment during total knee replacement.

DateNov 12, 2014 (initiated Sep 22, 2014)
StatusTerminated
ReasonMedacta has registered in its database 3 cases of breakage of the GMK Intramedullary Extension Rod 100 mm, code 02.07.10.0162.
Root causeNonconforming Material/Component
ActionMedacta sent an URGENT MEDICAL DEVICE letter dated September 22, 2014 to the sole US consignee. The letter included instructions for the consignee to: 1) check its inventory of the device; 2) quarantine and do not use the recalled device; and, 3) return the recalled device upon receipt of a RMA # from the Medacta QA Department. For questions regarding this recall call 805-910-6511.
Quantity1 rod (US); 145 rods (global)
DistributionWorldwide Distribution - US including OH and Internationally to Australia, Austria, Belgium, France, Germany, Greece, Italy, Japan, Portugal, South Africa, Spain, and Switzerland.