ClearanceBrief
Recall Z-0854-2015 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Artis zee/ zeego systems, model numbers: 10094135, 10094137, 10094139, 10094141, 10280959: dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipur

DateJan 14, 2015 (initiated Sep 17, 2014)
StatusTerminated
ReasonThe Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed, resulting in a short interruption of the procedure.
Root causeProcess control
ActionPlanned action (s) to repair defect or to bring product into compliance for Siemens Medical Solutions Inc. concerning the corrective action plan (CAP) for the Model(s) Artis Zee/Zeego Stationary Angiographic X-ray Systems. 1.Notifying the customer of the defect. 2.Implementing a corrective software patch to protect the system from damage in the event of hardware failure 3. Scheduling a service visit from a Siemens representative to perform a hardware replacement. 4. Replacing the non-conforming X-ray tube with one manufactured after the nonconformity issue was resolved. For further questi
Quantity44
DistributionUSA Distribution to the states of MO, MI, NY, OK, CA, LA, CT, IN, VA, MD, WA, HI, ME, TX, NY, OH, PA and CO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073290ARTIS ZEE AND ZEEGO
Siemens Medical Solutions USA, Inc., Feb 11, 2008