Recall Z-0192-2015 ยท OWB Interventional Fluoroscopic X-Ray System
GE OEC Medical Systems, Inc
OEC UroView 2800. The UroView 2800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical. and interventional procedures
| Date | Nov 7, 2014 (initiated Oct 3, 2014) |
| Status | Terminated |
| Reason | GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems. |
| Root cause | Component design/selection |
| Action | On October 3. 2014. GE Healthcare Surgery began distributing an Urgent Medical Device Correction letter to customers to inform them of the issue and provide instructions for immediate mitigation of the issue. The letter was sent to 3 titles at each facility affected: 1.1 Facility Administrator. 2.1 Director/Manager of Radiology and Radiologists, 3.) Radiology Department. Firm will track all service requests for the field action to completion. These service requests will document the inspection and component replacement (if applicable) on the affected units. Exceptions to completion of the serv |
| Quantity | 2 |
| Distribution | Worldwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K993687 | UROVIEW 2800 Ge Dec Medical Systems, Dec 21, 1999 |