Recall Z-0163-2015 ยท FFL Dislodger, Stone, Basket, Ureteral, Metal

C.R. Bard, Inc.

Bard Dimension Stone Basket Product Usage: The Bard Dimension Stone basket is a teardrop-shaped basket with the ability to capture stones by, opening and closing the basket and articulating or moving the basket from side to side. The device consists of 3 main parts; handle, shaft and basket. It is intended to use in the endoscopic removal of renal and ureteral stones.

DateOct 31, 2014 (initiated Oct 7, 2014)
StatusTerminated
ReasonThe product is non-sterile.
Root causeRelease of Material/Component prior to receiving test results
ActionBard sent an Urgent: Medical Device Recall letter to each customer on 10/7/2014 with proof of delivery notice. Wholesalers were asked to conduct a sub-recall. Letter identify the affected product, problem and actions to be taken. Customers were instructed to examine their inventory and quarantine product subject to recall, do not use or further distribute any of the affected product, remove and return affected product to Bard Medical Division, and if the product have been further distributed identify your customers and notify them at once of the product recall. Customers were instructed to
Quantity86 units
DistributionWorldwide Distribution in the states of AL, FL, NJ, NY, NC, PA, SC, TX, WI, and countries of Germany, Italy, Kuwait, and Luxembourg.