Recall Z-0174-2015 ยท DZE Implant, Endosseous, Root-Form

Straumann USA, LLC

Straumann Bone Level Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021.4308 Product Usage: Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery. Dental implants are intended for immediate pl

DateNov 6, 2014 (initiated Sep 24, 2014)
StatusTerminated
ReasonProduct transfer piece was fitted with a Narrow Connect (NC) instead of a Regular Connect (RC)
Root causePackaging process control
ActionStraumann USA sent a Field Corrective Action Notice letter dated September 24, 2014 and delivered via United Parcel Service. The letter identified the affected product, problem and actions to be taken. Straumann is requesting the return of unused implants. Returns and Customer Confirmation form is to be returned. Contact firm 781-747- 2514.
Quantity111 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121131BL, 04.1 MM RC, SLACTIVE 8MM, TIZR AND 10MM, 12, 14MM
Straumann USA, Jun 6, 2012