Recall Z-0129-2015 ยท DBF Ferritin, Antigen, Antiserum, Control

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur Calibrator C-Used with the ADVIA Centaur Ferritin assay on the ADVIA Centaur CP system only. (In-Vitro Diagnostic) Catalog Number: 06567787 (6 pack)

DateOct 24, 2014 (initiated Oct 9, 2012)
StatusTerminated
ReasonFerritin Analytical Sensitivity and Calibrator C Realignment with kit lots ending in 22 and 23 used with the ADVIA Centaur Ferritin assay on the ADVIA Centaur CP system
Root causeNonconforming Material/Component
ActionSiemens issued an Urgent Field Safety Notice via e-mail to all US customers affected on October 9, 2012. This notice informs the customer of the Field Correction and the actions that need to be taken. The Urgent Field Safety Notice instructs customers that upon receipt of Calibrator C Lots ending in 25 and higher (Cal C lots CC25LA and higher and Cal C lots CC25HA and higher), discontinue use of the Calibrator C lots ending in 22 and 23 with the ADVIA Centaur Ferritin assay when used on the ADVIA Centaur CP system. A confirmation fax-back form is included to ensure that all customers worldwide
Quantity2440 kits
DistributionWorldwide Distribution - Nationwide Distribution and the countries of Argentina, Australia, Brazil, Canada, China, Columbia, Egypt, Hong Kong, India, Israel, Indoenesia, India, Japan, Korea, Mexico, Malaysia, Philippines, Saudi Arabia, Singapore, Taipei, Uruguay, Venezuela, Viet Nam and South Afri
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110736ADVIA CHEMISTRY FERRITIN (FRT) METHOD
Siemens Healthcare Diagnostics, Inc., Aug 17, 2011