Recall Z-0152-2015 ยท IWB System, Radiation Therapy, Radionuclide

Elekta, Inc.

Laksell Gamma Knife Perfexion

DateOct 30, 2014 (initiated Oct 6, 2014)
StatusTerminated
ReasonElekta has become aware that the latches may be locked before they have been fully turned resulting in a poorly locked frame.
Root causeDevice Design
ActionElekta sent an" Important Field Safety Notice" distributed on October 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. This Important Field Safety Notice describes the problem and possible work arounds for customers to follow until a permanent solution can be developed. They are: " Always assure that the plastic lever is only operated when at a right angle with the frame adapter. " Do not force the plastic lever in place if it meets significant resistance when turned. " Ensure that the plastic lever is co
Quantity181
DistributionWorldwide Distribution - USA (nationwide) including the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OR, PA, RI, SC, TX, TN, UT, VA, WA and WV, and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120811LEKSELL GAMMA KNIFE PERFEXION
Elekta Instrument AB, May 24, 2012