ClearanceBrief
Recall Z-0936-2015 ยท IKK System, Isokinetic Testing And Evaluation

MedX Holdings, Inc.

Cervical Extension Machine for Physical Therapy

DateJan 7, 2015 (initiated Dec 3, 2013)
StatusTerminated
ReasonReview of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing.
Root causeUnknown/Undetermined by firm
ActionMedx began contacting its consignees December 30, 2013, to schedule testing on the devices at their facility. The technician has been scheduled to complete required testing of these devices in the field. For questions regarding this recall call 322-351-2005.
Quantity4
DistributionWorldwide Distribution - USA including MO, IL, ID, and MN and Internationally to Germany and South Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K896553MEDX CERVICAL EXTENSION TEST & REHAB MACHINE
Medx Corp., Jan 2, 1990