Recall Z-0166-2015 ยท MEA Pump, Infusion, Pca

Smiths Medical ASD, Inc.

CADD-Solis Medication Safety Software, Administrator CD, Version 3.1, REF 21-2194-0301-01. Product Usage: The CADD-Solis Medication Safety Software- Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD-Solis VIP Ambulatory Infusion Pump, CADD-Solis Ambulatory Infusion Pump, or CADD-Prizm PCSII Ambulatory Infusion Pump (software revision H or

DateNov 1, 2014 (initiated Sep 29, 2014)
StatusTerminated
ReasonSmiths Medical has identified an issue with a single batch (Lot Number 2752712) of CADD"-Solis Medication Safety Software Administrator 3.1 CDs. CADD"-Solis Medication Safety Software Administrator 3.1 CDs were sent with the CADD"-Solis Medication Safety Software Point of Care 3.1 software loaded on them.
Root causeProcess control
ActionConsignees were hand delivered by Smiths Medical Sales representatives on 9/29/2014 a Smiths Medical "Urgent Field Safety Notice" dated 25-September-14. The letter described the problem and the product being recalled. The letter also provided the Advice on Action to be Taken by the User, Transmission of this Urgent Medical Device Recall Notice and requested consignees to complete and return the Confirmation Form by fax to 1-800-237-8033 or via email to recall.response@smiths-medical.com. For questions customers can contact Smiths Medical's Customer Service Department at 1-800-258-5361.
Quantity24
DistributionUS Nationwide Distribution in the state of AK, CA, CO, CT, GA, MI, MN, NH, NC, SC, TX, VA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072144CADD -SOLIS AMBULATORY INFUSION PUMP AND CADD -SOLIS MEDICATION SAFETY SOFTWARE - ADMINISTRATOR
Smiths Medical MD, Inc., Mar 7, 2008