Recall Z-0151-2015 ยท FNL Bed, Ac-Powered Adjustable Hospital

Joerns Healthcare

Joerns EasyCare Bed Product Usage: The AC-powered adjustable hospital bed is a device intended for medical purposes. The bed includes a built-in electric motor and remote controls that can be operated by the patient to adjust the height and surface contour of the bed.

DateOct 30, 2014 (initiated Sep 24, 2014)
StatusTerminated
ReasonThe Easy-Care Care-Lock feature installed at the foot of the bed may potentially fail to stay engaged (locked) after the Care-Lock System has been engaged by stepping on the red (locked) pedal.
Root causeDevice Design
ActionJoerns sent an Urgent: Medical Device Correction letter dated September 24, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. The firm will provide the customers with a temporary solution kit to safegard the bed until a replacement Care-Lock assembly can be shipped and installed. Customers were instructed to complete the response form. For questions call 800-826-0270.
Quantity3135
DistributionWorldwide Distribution: US (nationwide) and country of: Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K951735JOERNS HEALTHCARE, INC. FULL LENGTH TELESCOPING RAIL
Joerns Healthcare, Inc., Aug 18, 1995