Recall Z-0133-2015 ยท DXC Clamp, Vascular

Sam Medical Products

SAM Junctional Tourniquet (SJT) Auxiliary, Model SJT 102 and SJT 112. The SJT models specific intended uses are to control difficult bleeds in the inguinal area and to immobilize a pelvic fracture.

DateDec 16, 2014 (initiated Aug 14, 2014)
StatusTerminated
ReasonSAM Junctional Tourniquet (SJT) Auxiliary strap, model SJT 102 and SJT 112, is recalled because the snap hook on auxiliary strap broke during training demonstration.
Root causeComponent design/selection
ActionSAM Medical Products issued the press release on August 29, 2014 and the link is http://www.ireachcontent.com/news-releases/sam-medical-products-implements-voluntary-recall-of-its-accessory-axilla-strap-for-the-sam-junctional-tourniquet-273239381.html. SAM Medical Products sent the "URGENT: MEDICAL DEVICE RECALL" letter, dated September 10, 2014, to their consignees. The 5 additional lot numbers (7/1/14, 7/2/14, 8/4/14, 8/7/14, and 8/11/14) that were erroneously placed on the recall notification letter and they are NOT part of the recall. SAM Medical Distributors will need to notify their
Quantitytotal 3633 units (3342 units in US and 291 units Outside US)
DistributionWorldwide Distribution: US (nationwide) in states of Ohio, Oregon, New Mexico, North Carolina, Texas and to the following countries: Australia, Canada, Hong Kong, Iceland, Japan, Lebanon, Singapore, Thailand, and United Kingdom.