Recall Z-0004-2015 ยท KPR System, X-Ray, Stationary

Siemens Medical Solutions USA, Inc

Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedur

DateOct 2, 2014 (initiated Sep 2, 2014)
StatusTerminated
ReasonTwo malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
Root causeSoftware design
ActionSiemens sent a Safety Advisory Notice, dated September 2, 2014, to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disable the automatic RIS worklist update and perform this procedure manually. Customers should disable the Measurement Field Display and Collimator Display by setting the display time for both setting to 0 sec (if editable). Customers should forward the safety notice to the new owner and inform Siemens of the new owner where possible. Customers with questions should call 610-21
Quantity2
DistributionNationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101491UROSKOP OMNIA
Siemens Medical Solutions USA, Inc., Jun 30, 2010