Recall Z-0093-2015 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Mckesson Information Solutions LLC

McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.

DateOct 20, 2014 (initiated Jun 25, 2014)
StatusTerminated
ReasonMcKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i
Root causeLabeling design
ActionConsignees were notified via Field Safety Notice on 06/25/2014.
Quantity625 devices
DistributionWorldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131497MCKESSON CARDIOLOGY HEMO
Mckesson Israel , Ltd., Dec 5, 2013