ClearanceBrief
Recall Z-0117-2015 ยท DZE Implant, Endosseous, Root-Form

Straumann USA, LLC

Straumann Bone Level Tapered Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021 .5308 Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery.

DateOct 23, 2014 (initiated Sep 19, 2014)
StatusTerminated
ReasonAssembled with Narrow Connect (NC) transfer piece instead of Regular Connect (RC)
Root causePackaging process control
ActionStrauman notified consignees of this Field Corrective Action by letter dated September 19, 2014 and delivered via hand by Straumann Representative. Accounts are requested to return inventory. Contact (978) 747-2514 for assistance.
Quantity6 units
DistributionUS Distribution in states of: FL and MA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140878STRAUMANN BONE LEVEL TAPERED IMPLANT
Straumann USA, LLC, Aug 20, 2014