Recall Z-0042-2015 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Mckesson Information Solutions LLC

McKesson Cardiology" Hemo intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.

DateOct 8, 2014 (initiated Aug 27, 2014)
StatusTerminated
ReasonSoftware Error: The McKesson Cardiology Hemo calculation section incorrectly converts the Hemoglobin value before it is utilized in the applicable formula calculations.
Root causeSoftware design (manufacturing process)
ActionOn August 27, 2014, McKesson initiated a correction via a Field Safety Notice to all affected domestic customers. McKesson Israel will be responsible for contacting foreign consignees.
Quantity272 devices
DistributionWorldwide distribution: US (nationwide) including states of: AR, CA, FL, DE, GA, HI, IL, IN, KS, LA, MA, MD, MN, MO, MS, MT, NJ, NV, OK, PA, SC, TNTX, , UT, WA; and countries of: Australia, Canada, Israel and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131497MCKESSON CARDIOLOGY HEMO
Mckesson Israel , Ltd., Dec 5, 2013