Recall Z-0086-2015 ยท MUE Full Field Digital, System, X-Ray, Mammographic

Siemens Medical Solutions USA, Inc

Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

DateOct 17, 2014 (initiated Sep 18, 2014)
StatusTerminated
ReasonThe stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i
Root causeProcess control
ActionA safety advisory notice, dated September 18, 2014, was sent to end users that identified the product, problem, and actions to be taken.
Quantity3
DistributionUS Distribution in states of: NY, VA, and MA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122286MAMMOMAL INSPIRATION
Siemens Medical Solutions USA, Inc., Feb 22, 2013