Recall Z-0149-2015 ยท JQW Station, Pipetting And Diluting, For Clinical Use

Tecan US, Inc.

Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.

DateOct 29, 2014 (initiated Sep 26, 2014)
StatusTerminated
ReasonDistortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.
Root causeNonconforming Material/Component
ActionTecan sent an Urgent Field Corrective Action letter dated September 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Tecan has instructed field engineers to replace the ZAAP Motion Board in the Freedom EVO instrument. Your instrument was identified as affected and a Tecan Service Representative will contact you to arrange replacement of the board. Please make sure that your internal controls would detect any possible dispense errors that might affect your application when using the instrument before replacemen
Quantity2 units
DistributionUS Distribution to IL.
FDA recordOpen on fda.gov