ClearanceBrief
Recall Z-0423-2015 ยท HWA Impactor

Zimmer, Inc.

Trabecular Metal Tibial Impactor. Product Usage: TM Tibial Impactor pads are used in conjunction with the TM Tibial Impactor to impact the tibial tray onto the tibia.

DateNov 24, 2014 (initiated Oct 14, 2014)
StatusTerminated
ReasonTo provide clarifying instructions relating to the assembly/disassembly of the Trabecular Metal Tibial Impactor with/from TM tibial Impactor Pads due to complaints regarding Tibal Impactor pad fracture.
Root causeLabeling design
ActionOn 10/14/2014, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected distributors and hospital Risk Managers. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-877-946-2761.
Quantity1165 units
DistributionWorldwide Distribution - US Nationwide in the state of AL, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, ND, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, WI, NC, TN, OK, AZ, UT, AK, PR, DC. and the countries of: Germany, China, Canada, Australia, Japan, United Arab Emirates, Austria
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.