Recall Z-0105-2015 ยท HGM System, Monitoring, Perinatal

Philips Medical Systems, Inc.

Philips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705A Intended for: " Monitoring the physiological parameters of pregnant women " Non-invasive monitoring of fetal heart rates and movements.

DateOct 21, 2014 (initiated Oct 3, 2014)
StatusTerminated
ReasonPhilips Intellivue and Avalon Fetal Monitors in time-synchronized automatic/sequence mode, the NBP automatic measurement series is stopped
Root causeSoftware Design Change
ActionPhilips Healthcare issued on 10/3/14 the Urgent Medical Device Correction notification/Field Safety Notice. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of a software upgrade free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for service. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medic
Quantity118 units
DistributionWorldwide Distribution-USA (nationwide) and the countries of AUSTRALIA AUSTRIA BAHRAIN BELGIUM CHINA COLOMBIA CZECH REPUBLIC DENMARK FINLAND FRANCE GERMANY HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN KAZAKHSTAN NETHERLANDS NEW ZEALAND NICARAGUA NIGERIA NORWAY OMAN POLAND ROMANIA RUSSIA SINGAPORE SLO
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111083PHILIPS AVALON FETAL/MATERNAL MONITORS FM20, FM30, FM40 AND FM50
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, Aug 17, 2011