ClearanceBrief
Recall Z-0098-2015 ยท KDQ Bottle, Collection, Vacuum

DeRoyal Industries Inc

DeRoyal(R) SafeLiner Suction Canister Kit, SEMI-RIGID, 1000cc, REF 71-6518, NON-STERILE, Rx Only, general hospital

DateOct 20, 2014 (initiated Aug 22, 2014)
StatusTerminated
ReasonThese lots of suction canisters and lids have an insufficient seal under low or intermittent suction condition.
Root causeProcess control
ActionDeRoyal sent an Urgent Voluntary Recall letter dated August 22, 2014, to all affected consignees by overnight letter or by email. DeRoyal was to follow up with non-responding accounts before October 1, 2014. Customers were instructed to check their inventory for the affected product. If they have affected product, customers were instructed to destory both the liner and the lid and dispose of it according to their facility's procedures. Customers were instructed to return the included Notice of Destruction to recalls@deroyal.com even if they no longer have affected product. Customers with
Quantity5112 units
DistributionNationwide Distribution including MI, OH, IN, FL, WV, VA, MI, NC, and NY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.