Recall Z-0008-2015 ยท CBK Ventilator, Continuous, Facility Use

Respironics California Inc

Esprit V1000 and V200 Ventilators, Model No. V1000 and V200; and the 3rd Generation Power Supply Repair Part (RP)Kits Microprocessor-controlled, electrically powered mechanical ventilators intended for use by qualified medical personnel to provide continuous or intermittent ventilator support for adult, pediatric and neonatal patients as prescribed by a physician.

DateOct 21, 2014 (initiated Sep 17, 2014)
StatusTerminated
ReasonA failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the ventilators will not operate. This failure will cause the loss of ventilator support, which may result in hypercarbia or hypoxemia.
Root causeUnknown/Undetermined by firm
ActionPhilips Healthcare sent an Urgent - Field Safety Notice dated September 17, 2014, to all customers who purchased the Esprit V1000 and V200 Ventilators to inform them that Respironics is recalling Esprit V1000 and V200 Ventilators installed with 3rd Generation Power Supplies and also 3rd Generation Power Supply Repair Part Kits. The letter identifies the affected devices, informs the customers of the problems, hazard involved, and the actions to be taken. Customers were reminded to be advised of the warnings provided. A Philips Respironics representative will contact customers to schedule a
Quantity2,206 units total (665 units in US)
DistributionWorldwide Distribution - USA, Argentina, China, Egypt, Ethiopia, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Jordan, Libyan Arab Yamahiriya, Malaysia, Mali, Mexico, Myanmar (Burma). Nepal, Nigeria. Pakistan, Peru, Philippines, Serbia, Singapore, South Africa, Sri Lanka, United Repu
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001208ESPIRT
Respironics, Inc., May 12, 2000
K023350ESPRIT VENTILATOR WITH RESPIRATORY MECHANICS
Respironics California, Inc., Nov 6, 2002
K034040ESPRIT VENTILATOR, MODEL V1000
Respironics, Inc., Jan 21, 2004
K981072ESPRIT VENTILATOR
Respironics, Inc., Nov 12, 1998