Recall Z-0948-2015 ยท JJH Clinical Sample Concentrator

Abbott Molecular

The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing.

DateJan 9, 2015 (initiated Nov 23, 2013)
StatusTerminated
ReasonThe x-drive assembly may fail prematurely. This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet.
Root causeNonconforming Material/Component
ActionAbbott Molecular sent a different Field Correction Recall / Urgent Field Safety Notice letters dated October 24, 2014 to each of three sets of customers. Customers whose instruments' X-Drive Assembly belts were perviously inspected by a Field Service Representative and found to not require replacement were advised to contact their local Abbott Molecular Representative or Abbott Molecular Technical Support at 800-553-7042 x 2 should they hear any unusual noises associated with the LiHa or RoMa movement or misalignment of the LiHa or RoMa. Customers whose instruments' X-Drive Assembly belts
Quantity176 devices
DistributionWorldwide Distribution - USA including CA, FL, GA, IL, MI, NC, NJ, NY, OH, OK, OR, and TX and Internationally to Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, China, Colombia, Costa Rica, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Indonesia, Ireland, Israel,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092705ABBOTT M2000SP AND ABBOTT M2000RT, MODELS 9K14-01 (G-SERIES), 9K14-02 (E-SERIES), 9K1501
Abbott Molecular, Inc., May 28, 2010