Recall Z-0103-2015 ยท LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur Systems HBc Total (HBcT) Kit. For In Vitro Diagnostic Use. Catalog Number: 07566733 Siemens Material Number: 10309508 The ADVIA Centaur HBc Total assay is an in vitro diagnostic test for the qualitative determination of total antibodies to the core antigen of the hepatitis B virus (HBc Total) in human serum or plasma (potassium EDT A, or lithium or sodium heparinized) using the ADVI

DateOct 21, 2014 (initiated Sep 15, 2014)
StatusTerminated
ReasonIncreased imprecision and positive bias is observed with patient samples and negative controls and can result in an increase in initial reactive results.
Root causeUnknown/Undetermined by firm
ActionSiemens Healthcare issued An Urgent Medical Device Recall (UMDR) sent to all affected Siemens Healthcare Diagnostics customers in the United States on September 15, 2014. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on September 15, 2014. These notices inform the customer of the imprecision and positive bias and a potential increase in initial reactive results associated with ADVIA Centaur Systems HBc Total Lot 100064. Customers were instructed to discontinue use of the affected lots and dispose of the product according to local
Quantity5434
DistributionWorldwide Distribution-USA (nationwide) and the countries of Austria Bahrain Belgium Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iraq Ireland Italy Jordan Latvia Lebanon Lithuania Luxembourg Netherlands Norway Poland Portugal Romania Slovakia Sp
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.