Recall Z-0451-2015 ยท IZI System, X-Ray, Angiographic

GE Healthcare, LLC

Innova IGS 540 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.

DateDec 16, 2014 (initiated Sep 4, 2014)
StatusTerminated
ReasonLoss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
Root causeDevice Design
ActionConsignees were sent on 9/4/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref#12218 dated September 4, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. Customers are to perform a system functional check as instructed in the letter. If the check fails, they should contact their local GE Healthcare Service Representative. GE Healthcare will correct all systems at no cost to t
Quantity10 (4 US, 6 OUS).
DistributionWorldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122457GE INNOVA/INNOVA IGS/DISCOVERY IGS/OPTIMA ANGIOGRAPHIC, FLUOSCOPIC X-RAY SYSTEMS WITH CATHLAB FRONTIERS SOLUTIONS
GE Healthcare, Jan 2, 2013