Recall Z-0214-2015 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur and ADVIA Centaur XP and ADVIA Centaur CP Cleaning Solution (For in vitro diagnostic use with the ADVIA Centaur systems) Catalog Number: 112748 (Siemens Material Number (SMN): 10310041 The Cleaning Solution is a component of the ADVIA Centaur and ADVIA Centaur XP and ADVIA Centaur CP instruments, and therefore does not have independent Instructions for Use (IFU). Information regard

DateNov 14, 2014 (initiated Sep 12, 2014)
StatusTerminated
ReasonStability of the ADVIA Centaur Cleaning Solution is not assured
Root causeMaterial/Component Contamination
ActionSiemens issued an Urgent Medical Device Recall (UMDR) to customers in the US on September 12, 2014. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on September 12, 2014. These notices inform the customer of the degradation of hypochlorite in these lots of cleaning solution. Customers were instructed to discontinue use of the affected lots and dispose of the product according to local regulations or procedures. The notices should be reviewed with the facility's laboratory or medical director. Fax-back forms will be used to confirm
Quantity4645 (12 pk)
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada. Grerat Britain, Saudi Arabia, Nepal, and Viet Nam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.