ClearanceBrief
Recall Z-0090-2015 ยท JXG Shunt, Central Nervous System And Components

Codman & Shurtleff, Inc.

Codman EDS 3 Drainage System (no Ventricular Catheter) (External Drainage System) Catalog Number: 82-1731 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated

DateOct 20, 2014 (initiated Sep 15, 2014)
StatusTerminated
Reasontubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)
Root causeDevice Design
ActionCodman Neuro sent an Urgent Notice dated September 12, 2014, to all affected customers via overnight mailing and OUS customers notified via email the week of September 15, 2014. Accounts were requested to isolate all inventory of the affected product and return to Stericycle at the following address: Stericycle 2670 Executive Drive, Suite A Attn: Event #8894 Indianapolis, IN 46241 For patients currently being managed with the EDS 3 System, the system should be replaced immediately. In the case where no substitute drainage system is immediately available, the EDS 3 System may continue to be u
Quantity93,558 units
DistributionWorldwide Distribution - USA (nationwide) and Internationally to: United Arab Emirates Argentina Austria Australia Aruba Belgium Bahrain Bolivia Brazil Bahamas Canada Switzerland Chile People's Republic of China Columbia Costa Rica Cyprus Czech Republic Germany Algeria Estonia Egypt Spain Finland Fr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061568CODMAN EDS 3 CSF EXTERNAL DRAINAGE SYSTEM, MODELS 82-1730, 82-1731, 82-1732, 82-1733, 82-1735, 82-1738
Codman & Shurtleff, Inc., Sep 29, 2006