Recall Z-0060-2015 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

GE Healthcare Carescape Patient Data Module

DateOct 9, 2014 (initiated Sep 19, 2014)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue to the ECG calculations following a disconnect/reconnect cycle with the Patient Data Module, when used with the Carescape Bx50 monitors. If the PDM is quickly disconnected and reconnected (a less than 10 second cycle) from a Carescape Bx50 monitor following a Patient Discharge or during active monitoring, the ECG waveform an
Root causeSoftware design
ActionConsignees were sent on 9/19/2014 a GE Healthcare "Urgent Medical Device Correction" letter Ref# 336107 dated September 18, 2014. The letter was addressed to Director of Biomedical Engineering, Chief of Nursing and Healthcare Administrator / Risk Manager. The letter described the Safety Issue, Safety Instructions , Affected Product Details, Product Correction and Contact Information. For questions they can contact Technical Support at 1-800-558-70444 or your local Service Representative.
Quantity68,452 (39,505 units US, 24,711 units OUS, 4,236 units unidentified/unknown)
DistributionWorldwide Distribution: US (nationwide) including DC, PR, Guantanamo except DE, HI, ND, SD & WY; and countries of: SOUTH AFRICA, KOREA, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TUNISIA, TRINIDAD TOBAGO, THAILAND, TANZANIA, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SLOVAKIA, SINGAPORE, SAUDI ARABIA, R
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071073SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE
Ge Medical Systems Information Technologies, May 11, 2007