Recall Z-0024-2015 · JIT Calibrator, Secondary

Siemens Healthcare Diagnostics, Inc.

Dimension Vista Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnostic product for the calibration of Creatine Kinase MB Isoenzyme (MMB) method on the Dimension Vista System.

DateOct 6, 2014 (initiated Aug 28, 2014)
StatusTerminated
ReasonSiemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf life of this product. Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been d
Root causeProcess design
ActionAn Urgent Medical Device Recall notice, dated August 2014, was provided to all Dimension Vista MMB CAL customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers. Customers were instructed to discard lots 3JD045, 4BD064, 4CD012, and 4DD065, and recalibrate with lot 4FD073.
Quantity6562
DistributionWorldwided distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Canada, Denmark, Germany, Finland, France, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, Slovenia, South Korea, Spain, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051087DIMENSION VISTA INTEGRATED SYSTEM
Dade Behring, Inc., Jul 11, 2005