Recall Z-2680-2014 ยท KPR System, X-Ray, Stationary

Siemens Medical Solutions USA, Inc

The Ysio Systems with software version VC10 The Ysio enables radiographic and tomographic exposures of the whole Body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.

DateSep 23, 2014 (initiated Aug 27, 2014)
StatusTerminated
ReasonSiemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image. If these artifacts appear in the region of interest, the examination may need to be repeated.
Root causeUnder Investigation by firm
ActionThe firm, Siemens, sent a "Safety Advisory Notice" dated August 27, 2014, to end users/customers. The notice identified the product, problem, and action to be taken. The customers were instructed to follow the instructions in the notice; promptly notify and instruct all the staff at your organization who have to be aware of this problem; forward this safety information to other organizations affected by this measure; and if the device has been sold and therefore no longer in your possession, forward to new owner. Siemens prepared a field modification of the Ysio systems with a hardware upgrad
Quantity62
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081722YSIO
Siemens Medical Solutions USA, Inc., Aug 22, 2008