ClearanceBrief
Recall Z-2691-2014 ยท GDW Staple, Implantable

Justright Surgical, LLC

Justright Surgical, JustRight 5mm Reload, REFJR-REL25-2.0, Sterile, EO, Rx Only It is a replacement cartridge used when more than one firing is required during a surgery.

DateSep 23, 2014 (initiated Aug 27, 2014)
StatusTerminated
ReasonJustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.
Root causeProcess design
ActionConsignees were notified via letter on August 27, 2014 in addition, the sales rep will hand deliver the Recall Notification on 8/28/14. Acknowledgement Form will be provided and returned products will be sent to STD Med where it will be held in quarantine.
Quantity3 cases containing 36 units
DistributionDistributed in the states of Colorado, California and New York.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132472JUSTRIGHT 5MM STAPLER, JUSTRIGHT 5MM RELOAD
Justright Surgical, LLC, Feb 11, 2014