Recall Z-2571-2014 ยท ITX Transducer, Ultrasonic, Diagnostic

Siemens Medical Solutions USA, Inc.

V5Ms Transesophageal transducers; V5Ms trans-esophageal echocardiograph (TEE) ultrasound transducer is intended primarily for cardiology applications

DateSep 3, 2014 (initiated Aug 13, 2014)
StatusTerminated
ReasonReports of deterioration of material covering the articulating section of the V5Ms transesophageal transducer.
Root causeUnder Investigation by firm
ActionSiemens sent a Customer Safety Advisory notification to all affected customers on August 13, 2014. The notification identified the product, the problem, and the action to be taken by the customer. Customers were advised that if they notice any damage or wear on their transducer, or if the transducer fails to pass the leakage current test, immediately discontinue use of the transducer and contact their local Customer Service Engineer. If they have any questions, customers should contact their local service support person for information regarding timelines and status. Customers were instructe
Quantity321 units
DistributionWorldwide Distribution - USA (nationwide) and Internationally tAustralia, Austria, Belgium, Bolivia, Brazil, Canada, China, Columbia, France, Germany, Great Britain, India, Ireland, Italy, Japan, Mexico, Netherlands, Norway, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Slovak Republic,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121807V5MS TRANSDUCER
Siemens Medical Solutions USA, Inc., Jul 10, 2012