Recall Z-0325-2016 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Medical Systems, Inc.

Ingenia, Intera, Achieva and Achieve dStream MR systems on Software versions R5.1.7 and R5.1.8 with the Mobi View software option. Diagnostic imaging system.

DateNov 23, 2015 (initiated Aug 21, 2014)
StatusTerminated
ReasonWhen a fused series of a sagittal, coronal or radial multi station scan is generated in MobiView, the resultant image order is reversed. This occurs with software release R5.1.7/R5.1.8.
Root causeSoftware Design Change
ActionPhilips Healthcare sent an Customer Information Letter dated July 2014, on August 21, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips will issue a Field Change Order to install updated software in a Service Pack that makes the slice numbering consistent for fused and unfused images. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377
Quantity307 units
DistributionWorldwide Distribution - USA (nationwide) and to the countries of: Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, Denmark Estonia, Ethiopia. France, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Korea, Republic of, Libya, Malaysia, Mexico, Ne
FDA recordOpen on fda.gov