Recall Z-0913-2015 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

The Symbia Intevo Excel is a non-diagnostic SPECT ICT system with CT support for only attenuation correction and anatomical localization. The Symbia Intevo Series are xSPECT systems. These are SPECT and T systems integrated through xSPECT technology during image registration and reconstruction. This xSPECT integration enables the use of extra-modal information for high image quality and quantifi

DateJan 6, 2015 (initiated Aug 12, 2014)
StatusTerminated
ReasonSiemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.
Root causeDevice Design
ActionSiemens sent Customer Safety Advisory Notice letters to all customers of record. The letters instructed customers to: 1) continue to use their system(s); 2) place the safety advisory in the system's instructions for use; and, 3) to anticipate that their local service engineer will contact them to perform an inspection and schedule any necessary repairs by December 2014. Customers were also instructed to share the letter with their support staff and, if the equipment was sold or transferred, to forward the safety notice to the new owner of the equipment. Customers with questions can contact t
Quantity6 systems
DistributionWorldwide Distribution - USA (naitonwide) and Internationally to Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131634SYMBIA
Siemens Medical Solutions USA, Inc., Aug 16, 2013