ClearanceBrief
Recall Z-2699-2014 ยท BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Vapotherm, Inc.

Vapotherm Precision Flow, Disposable Patient Circuit PF-DPC-High. For use with the Precision Flow system Precision Flow" is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to blended medical air/oxygen mixture a

DateSep 25, 2014 (initiated Aug 13, 2014)
StatusTerminated
ReasonDisposable Patient Circuit that allows water to leak into the center gas lumen
Root causeProcess control
ActionVapotherm issued an Initial Notification on August 13, 2014. The letter was updated on August 20, 2014 to include two additional lots. The recall letter notified users that this defect is easily detectable during normal use and the probability of serious injury or death is remote. The letter was sent to raise awareness of the recent increase in incidence rate of this issue and reminded users of the requirement in the operating manual that requires that the cannula be connected to the patient only after the set temperature is reached. The letter also advised patient care personnel not to use, o
Quantity22828 units
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada, France, Swizerland, Korea, Sweden, Germany , Belgium, Spain, Mexico, Scotland, UAE, Australia, Israel, Hong Kong, Ireland, kuwait, and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072845PRECISION FLOW
Vapotherm, Inc., Jul 17, 2008