ClearanceBrief
Recall Z-0911-2015 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

The Siemens Symbia S series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by

DateJan 6, 2015 (initiated Aug 12, 2014)
StatusTerminated
ReasonSiemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.
Root causeDevice Design
ActionSiemens sent Customer Safety Advisory Notice letters to all customers of record. The letters instructed customers to: 1) continue to use their system(s); 2) place the safety advisory in the system's instructions for use; and, 3) to anticipate that their local service engineer will contact them to perform an inspection and schedule any necessary repairs by December 2014. Customers were also instructed to share the letter with their support staff and, if the equipment was sold or transferred, to forward the safety notice to the new owner of the equipment. Customers with questions can contact t
Quantity877 systems
DistributionWorldwide Distribution - USA (naitonwide) and Internationally to Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131634SYMBIA
Siemens Medical Solutions USA, Inc., Aug 16, 2013